Advanced Search

Journal Navigation

Journal Home

Subscriptions

Archive

Contact Us

Table of Contents

Sign In to gain access to subscriptions and/or personal tools.
Journal of Hand Surgery (British and European Volume)
This Article
Right arrow Full Text (PDF)
Right arrow References
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Saved Citations
Right arrow Download to citation manager
Right arrowRequest Permissions
Right arrow Request Reprints
Right arrow Add to My Marked Citations
Citing Articles
Right arrow Citing Articles via Google Scholar
Right arrow Citing Articles via Scopus
Google Scholar
Right arrow Articles by SIMPSON, N. S.
Right arrow Articles by KINNINMONTH, A. W. G.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by SIMPSON, N. S.
Right arrow Articles by KINNINMONTH, A. W. G.
Social Bookmarking
 Add to CiteULike   Add to Complore   Add to Connotea   Add to Del.icio.us   Add to Digg   Add to Reddit   Add to Technorati   Add to Twitter  
What's this?

Articles

External Fixation of the Distal Radius

A biomechanical study

N. S. SIMPSON
R. WILKINSON
J. C. BARBENEL
A. W. G. KINNINMONTH

From the Department of Bioengineering, University of Strathclyde and the Department of Orthopaedic Surgery, Glasgow Royal Infirmary, Glasgow, UK

Correspondence: N. S. Simpson, Orthopaedic Department, Raigmore Hospital, Inverness, Scotland, UK

External fixation is a useful method of treating unstable fractures of the distal radius. There is a lack of information regarding the behaviour of mini-fixation systems, particularly under cyclical loading, which would be expected to occur at the wrist.

This laboratory study was designed to investigate the mechanical characteristics of nine current fixation systems. A programme of loading was devised to mimic forces acting on the distal radius. Wooden dowel was used to mount the fixator. Distraction and compression forces were applied in an Instron rig. Displacement was monitored throughout the load programme.

No fixator failed at the loads tested. The small Hoffman rectangular frame proved stiffest in compression and extension, but was heavier than other devices.

A trend of progressive, permanent deformation became apparent in those fixators which contain plastic or composite materials. This feature has implications for loss of fracture reduction in the clinical setting.

Journal of Hand Surgery (British and European Volume), Vol. 19, No. 2, 188-192 (1994)
DOI: 10.1016/0266-7681(94)90161-9


Add to CiteULike CiteULike   Add to Complore Complore   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us   Add to Digg Digg   Add to Reddit Reddit   Add to Technorati Technorati   Add to Twitter Twitter    What's this?