| Sign In to gain access to subscriptions and/or personal tools. |
Mechanical Strength of Scaphoid FixationFrom the Department of Orthopaedic Surgery, University of Connecticut Health Center, Farmington, Connecticut, USA Correspondence: M.L. Newport, Department of Orthopaedic Surgery, University of Connecticut Health Center, 10 Talcott Notch Road, Farmington, Connecticut, USA 06034-4037. We have investigated five devices suitable for scaphoid fixation (ASIF 2.7 mm and 3.5 mm cannulated screws, Herbert, Herbert-Whipple, and Howmedica Universal Compression Screw). The biomechanical properties tested were compression and resistance to cantilever bending. There was no statistically significant difference in compression between devices. There were significant differences in resistance to cantilever bending, with the Howmedica screw being strongest in both failure mode and in ultimate failure strength.
Journal of Hand Surgery (British and European Volume), Vol. 21, No. 1,