An In Vitro Biomechanical Study Comparing the Spiral Linking Technique Against the Pulvertaft Weave for Tendon RepairFrom the Warwick Orthopaedics, Warwick Medical School, UK, the Centre for Orthopaedic Biomechanics, University of Bath, UK and the Royal United Hospital, Bath, UK Correspondence: Mr Yuri Igorevich Kulikov MRCSEd, Warwick Medical School Coventry, West Midlands, UK. Tel.: +44 247 696 8621; fax: +44 247 696 8617. E-mail:Yuri_Kulikov{at}yahoo.com A new spiral linking technique for tendon repair in which one end of the tendon is spiralled around the other end has been developed. Using pig trotter extensor tendons, the Pulvertaft weave technique was compared with this new technique. Twenty-five repairs using each technique were tested by tensile loading with an Instron testing machine. The spiral linking technique matched the strength of Pulvertaft method: the mean peak loads were 102 and 105 N, respectively. The Pulvertaft weave was stiffer than the spiral linking technique: mean stiffness of 11.1 and 6.7 N/mm, respectively. The spiral linking technique also absorbed considerably more energy: energy absorbed prior to failure to 90% of peak load, 1.75 and 1.13 kN mm, respectively. In conclusion, the spiral linking technique appears as strong as the Pulvertaft weave and we believe it is easier to perform.
Key Words: biomechanics suture techniques tendons tensile strength
Journal of Hand Surgery (European Volume), Vol. 32, No. 4,
377-381 (2007) |
||||